The instrument uses a large-screen true-color LCD as a display , friendly man-machine dialogue interface, vector diagram display and wiring judgment to check the correctness of the circuit sex provides a reliable basis. Full-touch conductive silicone keyboard operation mode, easy to operate and easy to learn. The instrument has built-in large-capacity data storage without loss of data after power failure, which can save the on-site verification data, and can store up to 1000 sets of on-site verification results. It can provide background computer management software, and transfer the results to the computer through U disk, and produce reports. , print, realize computerized management.
The multi-function handheld automatic ratio tester is easy to operate, complete in function, stable and reliable in data, and can also perform blind testing (accurately measure the ratio and group number without knowing any nameplate parameters of the transformer under test, and at the same time can Accurately measure the phase angle); it can meet the needs of various large, medium and small transformer ratio tests (including Z-type transformers, rectifier transformers, electric furnace transformers, non-whole-point phase-shift transformations, traction transformers, Scott and reverse Scotts transformer)
1. Variable ratio measurement range: 0.8~10000.
2. Fast measurement speed: complete the three-phase test within 1 minute.
3. Measurement accuracy:
Measurement accuracy of high voltage side voltage: 0.05%
Low voltage side voltage measurement accuracy: 0.1%
Phase angle measurement accuracy: 0.1°
Ratio measurement accuracy: 0.1% (0.8-3000)
0.2% (3000-10000)
(According to user requirements, full-scale 0.1-level equipment can be customized)
4. Weight: 3Kg