Wuhan Wugao Electric Measurement Co., Ltd

WDHG-LY4 Six-channel full-function power frequency transformer tester (Touch Screen)

It is designed for automatic testing of protection and metering CT/PT, suitable for laboratory and field testing.

• Supports detection of CT and PT

• Meet the requirements of GB1207, GB1208 and other regulations.

• No need to connect other equipments, a single machine can complete all the test items, and complete the excitation and ratio polarity test at the same time in one test.

• Comes with a miniature fast printer, which can directly print the test results on site.

• Easy to operate, with smart prompts, making it easier for users to get started. .

• Large-screen LCD, graphical display interface.

• Automatically give CT/PT (excitation) inflection point value according to regulations.

• Automatically give 5% and 10% error curves.

• It can save 3000 groups of test data, which will not be lost after power failure.

• Support U disk dump data, can be read by standard PC, and generate WORD report.

• Small and lightweight ≤28Kg, very useful for field testing.


Products Inquiry

Product Details

1651220628107467.jpg

Main test function

CT (protection, metering)

PT

• Volt-ampere characteristic (excitation characteristic) curve

• Volt-ampere characteristic (excitation characteristic) curve

• Automatically give inflection point value

• Automatically give inflection point value

• Automatically gives 5% and 10% error curves, and can measure the actual value of the limit coefficient (ALF)

• Ratio measurement

• Ratio measurement, polarity judgment

• Polarity judgment

• Ratio measurement

• Ratio measurement

• Phase (angular difference) measurement

• Phase (angular difference) measurement

• One pass flow test

• AC withstand voltage test

• AC withstand voltage test

• Secondary load test

• Secondary load test

• Secondary winding test

• Secondary winding test

• PT volt-ampere voltage multiple is adjustable (1.2—3.5)

•Measurable volt-ampere characteristics of medium and low voltage transformers and zero-sequence transformers


• Measurable GIS, casing CT


•Comprehensive test, complete excitation and ratio polarity in one test at the same time


Main technical parameters 

Item

Parameters

Working power

AC220V±10%, 50Hz

device output

02500V 5KVA Maximum setting value 20A

High current output

01000A

Excitation accuracy

≤0.5% (0.2%*reading+0.3%*range)

Secondary winding resistance measurement

Range

0.1300Ω

Accuracy

≤1.0% (0.4%*reading+0.6%*range)

Secondary actual load measurement

Range

5500VA

Accuracy

≤0.5% (0.2%*reading+0.3%*range)±0.1VA

Phase measurement (angular difference)

Accuracy

4min

Resolution

0.1min

CT ratio measurement

Range

≤50000A/5A (10000A/1A)

Accuracy

≤0.5% (0.2%*reading+0.3%*range)

PT ratio measurement

Range

≤500KV

Accuracy

≤0.5% (0.2%*reading+0.3%*range)

Working environment

Temperature: -10℃40℃, Humidity: ≤90%, Altitude: ≤1000m

Size and weight

Size: 450mm × 280mm × 340mm, Weight: ≤28Kg


发货.jpg




联系我们.jpg


Email:info@wgdchipot.com

Whatsapp: +86 13657267902

Skype:  sallyliao2006

Wechat:sallyliao ( 13657267902)

Tel:+86 13657267902

Address:13 / F, Building 2, Hangang Science and Technology Industrial Park, Jintan Road, Wuhan, Hubei, China.